The Testomat® 808 is a limit value monitoring device for water hardness and the further development of the Testomat® F-BOB.
 

 
Application

  • reverse osmosis plants
  • soft water for commercial purposes
  • pure water production plants
  • galvanization

 
 
 
Measuring range

Water hardness 0.02 – 3 °dH (0.4….54 ppm CaCO3)
 

Technical Data

Power supply: Depending on product configuration: 24V / 115V / 230V, 50 – 60 Hz
Instrument protection: 230 – 240 V: T0,1 A; 115 V: T0,2 A; 24 V: T0,8 A
Mains protection for consumers: max. 4 A (n , l)
Power consumption: max. 16 VA, without external load
Protection class: I
Degree of protection: IP 54
Conformity: EN 61000-6-2, EN 61000-6-4, EN 61010-1
Ambient temperature: 10 – 40 °C; 50 – 104°F
Current interface: Output of defined values (5, 8, 11, 14, 17, 20 mA) for displaying status and error messages, max. load 500 Ohms
Contact load relay: 230V / 4A AC ohm resistive load
Dimensions: B x H x T = 364 x 314 x 138 mm
Weight: 4350 g
 

Mains water supply

Operating pressure: Depending on product configuration:
0.3 – 1 bar or
1 – 4 bar
a pressure reducer (special accessory) should be used from 4 to 8 bar range
Water inlet:
Opaque hose with 6 mm external diameter/4 mm internal diameter
Water outlet: Hose with 6 mm external diameter/4 mm internal diameter
Water temperature: 10 to 40 °C

Here you will find available indicators for the Testomat® 808.
See accessories in the product search to find e.g. pumps, filters and service sets for the Testomat® 808.
 

Product video

This video shows how easily the Testomat® F-BOB can be replaced by the Testomat® 808.