The process photometer Testomat 2000 THCL® combines chlorine and hardness measurement in one device.
Application
- medical technology (dialysis)
- corrosion protection
- protection for reverse osmosis membranes
- monitoring of softener and chlorination systems for drinking water or swimming pools
Measuring range
mg/l (resolution) | °dH (resolution) | °f (resolution) | ppm CaCO3 (resolution) | mmol/l (resolution) | |
Total Chlorine |
0.0 – 2.5 (0.1) | ||||
Water Hardness | 0.25 – 2.50 (0.05) | 0.45 – 4.48 (0.10) | 4.48 – 44.8 (0.9) | 0.04 – 0.45 (0.01) |
Technical Data
Power supply: 230–240 VAC, 115 VAC, 24 VAC all 50–60Hz
Power consumption: max. 30 VA
Protection class: I
Degree of protection: IP 65
Ambient temperature: 10–45 °C, 50 – 104°F
Option: booster pump see accessories
Dimensions (B x H x T): 380 x 480 x 280 mm
Weight: approx. 9.5 kg
Menu languages: German, English, French, Italian
Mains water supply
Operating pressure:
1 to 8 bar (15 to 116 psi) or
0,3 to 1 bar (5 to 15 psi), after removing of the valve body out of the regulator/filter housing
Water inlet: Opaque hose with 6 mm external diameter/4 mm internal diameter
Water outlet: Hose with 6 mm external diameter/4 mm internal diameter
Water temperature: 10 to 40 °C; 50 to 104°F
Here you will find available reagents and available indicators for the Testomat 2000® THCL.
Select accessories in the product search to find e.g. pumps, filters and service sets for the Testomat 2000® THCL.