The Testomat® Limit TH process photometer automatically determines and monitors automatically the content of residual total hardness (water hardness) in water. It will provide you with a lot of new functions additionally to the approved functions of the Testomat 808® series which are: a TFT panel (touchscreen), integrated SD card, firmware update and Import and export of settings via SD card, optionally automatic programmable self-cleaning of the measuring chamber and many more.
Application
- Use in water treatment (e.g. osmosis plants, electroplating, large boiler plants, laundries, canteen kitchens, water softening systems)
Measuring range
0,05 – 25,0 °dH
The limit value to be monitored can be freely set within the measuring ranges of the Testomat 2000® indicators.
Technical Data
Power supply: 24V DC current; device protection T0,8 A
Power consumption: max. 16 VA; without external load
Contact load relay: 28VDC/300W or 45VAC/300W ohm resistive load
Current interface: Output of defined values (5, 8, 11, 14, 17, 20 mA) for displaying status and error messages, max. load 500 Ohms
Protection class: II
Degree of protection: IP 44
Ambient temperature: 10–40 °C, 50 – 104°F
Optional: automatic cleaning function after installation of a selfclean pump (see accessories)
Dimensions (B x H x T): 424 x 323 x 138 mm
Weight: approx. 4,5 kg
Mains water supply
Operating pressure:
1 to 8 bar (15 to 116 psi) or
0,3 to 1 bar (5 to 15 psi), after removing of the valve body out of the regulator/filter housing
Water inlet: Opaque hose with 6 mm external diameter/4 mm internal diameter
Water outlet: Hose with 12 mm internal diameter
Water temperature: 10 to 40 °C; 50 to 104°F
Here you will find available reagents for the Testomat® Limit TH.
Select accessories in the product search to find e.g. pumps, filters and service sets for the Testomat® Limit TH.