The Testomat® PRO FE process photometer automatically determines and monitors the the dissolved iron content (II)+(III) in water. It will provide you with a lot of new functions additionally to the approved functions of the Testomat series which are: integrated SD card, firmware update and Import and export of settings via SD card, optionally automatic programmable self-cleaning of the measuring chamber and many more.

 
Application

  • de-ironing plants
  • monitoring of process or drinking water flows

 
Measuring range

0 to 1 mg/l iron (II) + (III)
 

Technical Data

Power supply: multi voltage power supply unit 100-240 VAC +/- 10%, 50/60 Hz
Power consumption: max. 230 V / 4A ; without external load max. 230 V / 1A
Durability relay: DC: 8A at 30V or 0.28A at 250V; AC: 8A at 415V
Protection class: I
Degree of protection: IP 44
Ambient temperature: 10–40 °C, 50 – 104°F
Dimensions (B x H x T): 380 x 480 x 280 mm
Weight: approx. 9.0 kg
Menu languages: German, English, French, Dutch (others upon request)
 

Mains water supply

Operating pressure:
1 to 8 bar (15 to 116 psi) or
0,3 to 1 bar (5 to 15 psi), after removing of the valve body out of the regulator/filter housing
Water inlet: Opaque hose with 6 mm external diameter/4 mm internal diameter
Water outlet: Hose with 12 mm internal diameter
Water temperature: 10 to 40 °C; 50 to 104°F

Here you will find available reagents for the Testomat® PRO FE.
Select accessories in the product search to find e.g. pumps, filters and service sets for the Testomat® PRO FE.