The Testomat® 808 SiO2 has been designed for use in the sterilisation of hospitals. It complies with the EN 285:2006 standard. The device is a limit gauge that automatically monitors the level of SiO2 in the water. You have a choice between 10 limits in the range of 0.3 to 1.2 ppm.
Application
- sterilisation of hospitals
- silica monitoring
Measuring range
Silica SiO2: 0.3 – 1.2 ppm
Technical Data
Power supply: Depending on product configuration: 24V / 115V / 230V, 50 – 60 Hz
Instrument protection: 230 – 240 V: T0,1 A; 115 V: T0,2 A; 24 V: T0,8 A
Mains protection for consumers: max. 4 A (N , L)
Power consumption: max. 16 VA, without external load
Protection class: I
Degree of protection: IP 54
Conformity: EN 61000-6-2, EN 61000-6-4, EN 61010-1
Ambient temperature: 15 – 25 °C; 59 – 77°F
Current interface: Output of defined values (5, 8, 11, 14, 17, 20 mA) for displaying status and error messages, max. load 500 Ohms
Contact loadrelay: 230V / 4A AC ohm resistive load
Dimensions: B x H x T = 364 x 314 x 138 mm; with side pocket: 442 x 314 x 138 mm
Weight: 4350 g
Mains water supply
Operating pressure: Depending on product configuration:
0.3 – 1 bar or
1 – 4 bar
a pressure reducer (special accessory) should be used from 4 to 8 bar range
Water inlet: Opaque hose with 6 mm external diameter/4 mm internal diameter
Water outlet: Hose with 6 mm external diameter/4 mm internal diameter
Water temperature: 10 to 40 °C
Here you will find available indicators for the Testomat® 808 SiO2.
Select accessories in the product search to find e.g. pumps, filters and service sets for the Testomat® 808 SiO2.